First uncertainty evaluation of the FoCS-2 primary frequency standard
Author(s)
Date issued
2018
In
Metrologia, IOP
Vol
55
No
3
From page
366
To page
385
Subjects
atomic clocks primary frequency standards continuous atomic fountain caesium
Abstract
We report the uncertainty evaluation of the Swiss continuous primary frequency standard FoCS-2 (Fontaine Continue Suisse). Unlike other primary frequency standards which are working with clouds of cold atoms, this fountain uses a continuous beam of cold caesium atoms bringing a series of metrological advantages and specific techniques for the evaluation of the uncertainty budget. Recent improvements of FoCS-2 have made possible the evaluation of the frequency shifts and of their uncertainties in the order of 1 × 10<sup>−15</sup>. When operating in an optimal regime a relative frequency instability of 8 × 10<sup>−14</sup> (<i>τ</i>/<i>s</i>) <sup>−1/2</sup> is obtained. The relative standard uncertainty reported in this article, 1.99 × 10<sup>−15</sup>, is strongly dominated by the statistics of the frequency measurements.
Publication type
journal article
File(s)![Thumbnail Image]()
Loading...
Name
Jallageas_A.-First_uncertainty-20180716113243-ST.pdf
Type
Main Article
Size
5.19 MB
Format