First results with a cold cesium continuous fountain resonator

Author(s)
Dudle, Gregor
Joyet, Alain  
Laboratoire temps-fréquence  
Berthoud, Patrick
Mileti, Gaetano  
Laboratoire temps-fréquence  
Thomann, Pierre  
Institut de physique  
Date issued
2001
In
IEEE Transactions on Instrumentation and Measurement, Institute of Electrical and Electronics Engineers (IEEE), 2001/50/2/510-514
Subjects
Atomic fountain atomic frequency standards Dick effect
Abstract
We report on the design, construction, and preliminary measurements on the resonator of a continuous Cs fountain frequency standard. The construction of the resonator is described, preliminary measurements of the available atomic flux, and of the beam temperature are presented, along with the first Ramsey fringes (width ≃1 Hz) obtained in this new type of fountain. We discuss theoretical aspects of the interrogation scheme with a special view on how aliasing or intermodulation effects are suppressed in a continuous fountain.
Publication type
journal article
Identifiers
https://libra.unine.ch/handle/20.500.14713/57477
DOI
10.1109/19.918179
-
https://libra.unine.ch/handle/123456789/10574
File(s)
Loading...
Thumbnail Image
Download
Name

Dudle_Gregor_-_First_Results_with_a_Cold_Cesium_Continuous_Fountain_20120227.pdf

Type

Main Article

Size

1.06 MB

Format

Adobe PDF

Checksum

(MD5):dc026c61a95a95998e72c4a87dbe0432

Université de Neuchâtel logo

Service information scientifique & bibliothèques

Rue Emile-Argand 11

2000 Neuchâtel

contact.libra@unine.ch

Service informatique et télématique

Rue Emile-Argand 11

Bâtiment B, rez-de-chaussée

Powered by DSpace-CRIS

v2.0.0

© 2025 Université de Neuchâtel

Portal overviewUser guideOpen Access strategyOpen Access directive Research at UniNE Open Access ORCIDWhat's new