First uncertainty evaluation of the FoCS-2 primary frequency standard

2018, Jallageas, Antoine, Devenoges, Laurent, Petersen Mangin, Michael, Morel, J, Bernier, L. G, Schenker, Dominique, Thomann, Pierre, Südmeyer, Thomas

We report the uncertainty evaluation of the Swiss continuous primary frequency standard FoCS-2 (Fontaine Continue Suisse). Unlike other primary frequency standards which are working with clouds of cold atoms, this fountain uses a continuous beam of cold caesium atoms bringing a series of metrological advantages and specific techniques for the evaluation of the uncertainty budget. Recent improvements of FoCS-2 have made possible the evaluation of the frequency shifts and of their uncertainties in the order of 1 × 10−15. When operating in an optimal regime a relative frequency instability of 8 × 10−14 (τ/s) −1/2 is obtained. The relative standard uncertainty reported in this article, 1.99 × 10−15, is strongly dominated by the statistics of the frequency measurements.