Measurement of the magnetic field profile in the atomic fountain clock FoCS-2 using Zeeman spectroscopy
Author(s)
Date issued
2017
In
Metrologia, IOP
Vol
54
No
2
From page
239
To page
246
Subjects
atomic fountain clock frequency bias magnetometry Zeeman shift
Abstract
We report the evaluation of the second-order Zeeman shift in the continuous atomic fountain clock FoCS-2. Because of its continuous operation and geometrical constraints, the methods used in pulsed fountains are not applicable. We use here time-resolved Zeeman spectroscopy to probe the magnetic field profile in the clock. Pulses of ac magnetic excitation allow us to spatially resolve the Zeeman frequency and to evaluate the Zeeman shift with a relative uncertainty smaller than 5 × 10<sup>−16</sup>.