  4. Integrated Microanalytical System for Simultaneous Voltammetric Measurements of Free Metal Ion Concentrations in Natural Waters

Integrated Microanalytical System for Simultaneous Voltammetric Measurements of Free Metal Ion Concentrations in Natural Waters

Author(s)
Noël, Stéphane
Tercier-Waeber, Mary-Lou
Lin, Lin
Buffle, Jacques
Guenat, Olivier
Koudelka-Hep, Milena
Date issued
2006
In
Electroanalysis, Wiley, 2006/18/21/2061-2069
Subjects
Metal speciation Free metal ions Microsensor array Microanalytical system Square wave voltammetry Environmental analysis Natural waters
Abstract
A complexing gel integrated microelectrode (CGIME) for direct measurements of free metal ion concentrations in natural waters has been developed. It is prepared by the successive deposition of microlayers of a chelating resin, an antifouling agarose gel and Hg on a 100-interconnected Ir-based microelectrode array. The trace metals of interest are in a first step accumulated on the chelating resin in proportion to their free ion concentration in solution, then released in acidic solution and detected simultaneously by using square wave anodic stripping voltammetry (SWASV). The reliability of this sensor for the simultaneous measurement of copper, lead and cadmium has been studied by a series of replicate laboratory tests. The proportionality between the voltammetric peak current intensity and the free metal ion concentrations in solution has been demonstrated by using malonate as a model ligand. Finally, the CGIME sensor was applied to the Cu and Pb free concentration measurement in sea water samples and the results compared to the free metal ion concentrations measured using hollow fiber based permeation liquid membrane (HF-PLM) coupled to inductively coupled plasma mass spectrophotometer (ICP-MS). Comparable concentration values were found for both metals with both techniques allowing to validate the CGIME measurements in complex media.
Publication type
journal article
Identifiers
https://libra.unine.ch/handle/20.500.14713/57609
DOI
10.1002/elan.200503632
File(s)
Loading...
Thumbnail Image
Download
Name

No_l_St_phane_-_Integrated_Microanalytical_System_20071116.pdf

Type

Main Article

Size

467.56 KB

Format

Adobe PDF

Université de Neuchâtel logo

Service information scientifique & bibliothèques

Rue Emile-Argand 11

2000 Neuchâtel

contact.libra@unine.ch

Service informatique et télématique

Rue Emile-Argand 11

Bâtiment B, rez-de-chaussée

Powered by DSpace-CRIS

libra v2.1.0

© 2025 Université de Neuchâtel

Portal overviewUser guideOpen Access strategyOpen Access directive Research at UniNE Open Access ORCIDWhat's new